REF 15245 Incubator with an inner volume of 200 liters
Advantages of Labo C201:
7“ colour touch screen
UV decontamination (option)
Particle filter (P3) within airflow
Display of measured values for 2 h or 24 h (T, CO2, rH, O2)
Cable bushing (Access port)
Access control via numerical code (option)
RS485 or ethernet interface
Active sterile humidity:
External water reservoir
Relative humidity vaporized at 120 °C
Minimized risk of contamination
Humidity sensor (rH)
Very short recovery times
CO2 sensor:
Dual beam infrared sensor
Very precise measurement
Long-term stable
Direct heating:
Independent over-all heating of inner walls and door
Homogeneous temperature distribution in the interior
Optional:
UV decontamination
O2 control
Subdivided inner glass door with 6 small doors
Access control via numerical code
PC software LaboDat+
Features
Interior
Electro-polished stainless steel with coved corners Shelves and ductwork easily removable for cleaning purpose Cable bushing (Access port)
Temperature
Intelligent temperature control system for dry inner chamber
Humidity supply
Active sterile humidity via external water supply container
CO2 control
Dual beam infrared sensor Measurement independent from temperature and humidity
Short recovery times
For all adjustable parameters through optimized microprocessor control
Easy handling
Menu guided 7“ colour touch display Display of measured values for 2 h/24 h
Diagnostic system
Optical and acoustic alarm if set point deviation or sensor defect occurs Door monitoring Remote alarm
Options
O2 control: connection for N2 possible Subdivided inner glass door with 6 small doors Access control via numerical code UV decontamination PC software LaboDat+
Interfaces
Alarm contact (potential free) RS485 or Ethernet interface
Technical data
Exterior
Dimensions 720 x 700 x 870 mm (w x d x h)
Volume 200 l Electro-polished stainless steel Shelves height adjustable
Panel heated interior chamber and door Range 27 °C - 42 °C starting 5 °C above ambient temperature Stability / uniformity ± 0.1 °C / ± 0.3 °C Over-temperature alarm 32 °C - 47 °C
Dual beam infrared sensor Range 0 – 10 % CO2 Stability ± 0.1 % CO2
O2 control < 21 % (optional)
Measurement with galvanic sensor Connection of N2 Range 1 % – 21 % O2
Humidity
Active sterile humidity supply through vaporizing module operating at 120 °C Measuring range 0 – 98 % rH / range 60 % – 95 % rH
Access port
Ø 30 mm
Supply
230 V AC, 50/60 Hz or 115 V AC, 50/60 Hz, 250 VA Inlet pressure for gas 0.8 bar Ambient temperature 18 °C - 30 °C
Classification
Safety class I EN 61010 conform